EEG support
Up to 32 channels: configurable
2 references: CMS/DRL configurable in any 10-20 location or on ears (compatible ear clips for sensors included)
Available data: raw EEG and frequency band data. Requires PRO License for use.
EEG signals
Sampling method: Sequential sampling, single ADC
Sampling rate: 1024 internal downsampled to 128 SPS.
Resolution: 14 bits with 1 LSB = 0.51μV (16 bit ADC, 2 bits instrumental noise floor discarded), compressed maximum slew rate 32uV per sample.
Bandwidth: 0.16 – 43Hz, digital notch filters at 50Hz and 60Hz
Dynamic range (input referred): 8400 μV(pp)
Coupling mode: AC coupled